Mobile negative pressure isolation chamber
EST Company's Technical Background
Mr. Wang Chaofeng, the Director of EST, founded the China office of Imtech Environmental Simulation Testing Technology Company in Germany in 1998 and served as the Chief Representative General Manager.
Mr. Xu Jianchuan, the Chief Engineer of EST, joined Imtech in Germany in 2003 and has been engaged in the installation, commissioning, and after-sales service of imported environmental simulation test warehouses in Germany.
They inherited the environmental simulation testing system technology from Imtech, a German company. Since 2017, EST company has been carrying out the domestic transformation of Imtech environmental simulation test system technology. We have world-class and domestically leading technical experience in environmental simulation testing systems.
preface
After 23 years of working in scientific experimental environment test chamber technology, EST has developed civilian single person movable negative pressure isolation environmental chambers for different purposes based on the special needs of the current situation.
This environmental warehouse is used in airports, high-speed rail stations, docks, customs and other places.
The air pressure inside the small isolation environment chamber should always be lower than the surrounding environmental pressure, even including the instantaneous state of opening and closing the chamber door. The gas in the bin shall be strictly disinfection and sterilization by two levels of ultraviolet light to ensure the safety of surrounding personnel.
After passengers leave the cabin, the inside of the isolation cabin must be thoroughly disinfection and sterilization immediately.
This small isolated environmental compartment can also be equipped with a specially designed temperature regulation system as an option. Create a comfortable environment for temporary personnel in the warehouse.
This small isolated environmental compartment is equipped with a dedicated fresh air system, which means that the fresh air entering the compartment and the polluted air discharged will never mix with each other. Avoid the pollution risk caused by the mixing of exhaust and fresh air. The combination of various primary, intermediate, and high-efficiency filters in the system can ensure the biological cleanliness level of the environmental chamber.
In addition, the location of the small isolation environment warehouse is a public place. The exterior design must be integrated into the local installation environment as much as possible.
Also, the isolated personnel in this small isolation environment warehouse come from all over the world. The internal design should also strive to make isolated personnel feel warm.
system description
Warehouse body:
Warehouse dimensions: 2200× 1300× 2000 mm(H×W× D)
Internal dimensions of the warehouse: 2100 × 1200 × 1500 mm (H × W × D)
Outer warehouse version: sturdy steel plate, 1.5mm, spray painted, blue+milky white
Insulation layer: 50mm thick fireproof insulation material
Surrounding panels in the warehouse: milky white, or other light colors
Ceiling inside the warehouse: milky white, or other light colors
Floor: Non slip floor, 3mm, lined with composite reinforcement board
Personnel entry and exit: 1900× 850 mm
Circular observation window on the door: ϕ300mm, Double layered glass
Side: Double glass windows
Disinfection transfer window and observation window for items entering and exiting:
External dimensions: 400× 400 × 400 mm(H×W× D)
Internal available sizes: 300× 280 × 350 mm(H×W× D)
Internal and external window door interlock
Disinfection method: UV lamp
Lighting inside the warehouse: 200 LUX
Four movable casters that can be locked
Brand new air treatment and exhaust disinfection system
Fresh air enters from the air intake grille on the back of the compartment, passes through the G4 primary filter, and enters the fresh air modulation box.
The fresh air modulation box is installed on the back of the warehouse body
include:
G4 primary filter
Specially designed air temperature regulating unit circuit (optional)
Heater (optional)
Variable frequency fan
Regulating valve
Connected to the upper air supply duct behind the warehouse, fresh air is sent into the warehouse through the medium efficiency filter F7 and grille.
Specially designed cabin air temperature control unit (optional)
Cooling capacity: 1.5KW
Air output: 400M3/h
Air outlet temperature: minimum 5 ℃
The exhaust treatment box is also installed on the back of the warehouse
include:
Primary filter G4
Regulating valve
Variable frequency fan
First level UV disinfection
High efficiency filter T14
Secondary UV disinfection
(Please note that even for T14 high-efficiency filters, strict adherence to relevant regulations is required)
automatic control system
This system adopts Siemens' latest generation PLC, S7 as the controller. The PLC with Ethernet interconnection function collects signals from all sensors, drives various subsystems, and completes high-precision automation operation;
There are also emergency alarm buttons and intercom devices installed in the warehouse. Can add: camera, WIFI, etc.
System technical data
External environmental temperature: 5 ℃ to 35 ℃
Total fresh air volume: 70 to 200 m3/h, variable frequency adjustable
Warehouse pressure control:- 30Pa
Power Supply: AC 220V,50Hz, Three wire system
Total power consumption: 2000W
Including short-term power supply battery
Total weight: Approximately 250KG
We are currently debugging the isolation environment warehouse in the workshop and preparing it to leave the factory

Contact: Wang Chaofeng (13701191584)